Back to all papers

Query-based prediction of 3D x-ray scatter fields in an interventional suite using a transformer model.

October 6, 2026pubmed logopapers

Authors

Rettig N,Schülein P,Bukschat Y,Vetter M

Affiliations (1)

  • Institute for Applied Artificial Intelligence and Robotics (A2IR), Technical University of Applied Sciences Mannheim, Mannheim, Germany.

Abstract

Fluoroscopy-guided interventions expose clinical staff to x-ray scatter radiation, motivating accurate three-dimensional scatter-field predictions for radiation protection planning. Monte Carlo (MC) simulations provide reference accuracy but require minutes to hours of computation, while analytical models are faster but often lack spatial fidelity in complex geometries. We present a transformer-based model for scatter dosimetry in a single simulated interventional suite with fixed room geometry and object inventory, supporting near real-time inference on localised subvolumes and interactive runtimes for whole-room evaluation. We train a separate transformer-based model for each target quantity: air kerma per source photon, transported energy, and energy fluence per source photon. Each predictor encodes spectrum and technique parameters and represents room objects and their poses as a set of scene tokens. Predictions are obtained by coordinate-conditioned regression at queried voxel locations, so inference can be restricted to selected subvolumes and runtime scales with the number of queried voxels. The training set comprisessimulated configurations with varied C-arm poses, spectra, patient-table and patient phantom positions. On an independent test set ofconfigurations, the model achieves medianfor log-transformed targets and structural similarity around. In linear units, medianvalues arefor air kerma per source photon,for transported energy, andfor energy fluence. Full-room inference, reconstructing the effective grid of aboutmillion voxels, takeson an NVIDIA L40S GPU compared tofor MC simulation, yielding<i>a</i>speed-up. These results represent a step toward interactive, room-scale radiation protection planning, enabling rapid what-if analyses of staff positioning and procedure planning.

Topics

Scattering, RadiationImaging, Three-DimensionalJournal Article

Ready to Sharpen Your Edge?

Subscribe to join 11k+ peers who rely on RadAISlice. Get the essential weekly briefing that empowers you to navigate the future of radiology.

We respect your privacy. Unsubscribe at any time.