Quantitative and task-based performance of spectral deep learning reconstruction in low-dose dual-layer coronary CT angiography: A phantom study.
Authors
Affiliations (3)
Affiliations (3)
- Department of Radiology and Nuclear Medicine, University Medical Center Utrecht, Utrecht, Netherlands.
- Department of Radiology, University Medical Center Groningen, University of Groningen, Groningen, Netherlands.
- Clinical Science, Philips Healthcare, Best, Netherlands.
Abstract
Coronary CT angiography (CCTA) is an established noninvasive test for coronary artery disease, but balancing image quality and radiation dose remains challenging. Dual-layer spectral CT (DLCT) enables reconstruction of virtual monoenergetic images (VMI) that enhance coronary artery contrast, but are inherently more sensitive to noise, making this trade-off especially critical. Deep learning reconstruction (DLR) improves image quality and enables radiation dose reduction in CCTA, but its performance on spectral reconstructions from dual-layer CT (DLCT) remains unexplored. This phantom study compares spectral DLR at five different noise reduction strengths with hybrid-iterative reconstruction (hybrid-IR) at one fixed, clinically used noise reduction strength across conventional and VMI at multiple reduced dose levels. An anthropomorphic thorax phantom with a coronary insert containing two calcifications (100 and 800 mg CaHA/mL) was scanned on a DLCT system at six dose levels (volumetric CT dose index 12.2-1.4 mGy) using a clinical sequential CCTA protocol. Data were reconstructed with one clinical hybrid-IR protocol, and DLR at five noise reduction strengths (Smoother, Smooth, Standard, Sharp, Sharper) for conventional images and VMI at 7 energy levels (40-70 keV). Image noise, luminal contrast-to-noise ratio (CNR<sub>lumen</sub>), noise power spectrum means frequency (NPS<sub>mean</sub>), task transfer function at 50% (TTF<sub>50</sub>), and detectability index (d') of coronary plaques were assessed. Statistical significance was inferred when 95% confidence intervals did not overlap. At full dose, DLR significantly reduced noise up to 64% in conventional images and 67% in 40 keV VMI, while increasing CNR<sub>lumen</sub> up to 300% and 230%, respectively, compared with hybrid-IR. DLR Standard, Sharp, and Sharper significantly increased NPS<sub>mean</sub> by 7%-13% for conventional images and 40 keV VMI, whereas DLR Smoother shifted noise toward lower frequencies, reducing NPS<sub>mean</sub> by 22%-32%. All DLR strengths significantly improved spatial resolution of calcified plaque, with TTF<sub>50</sub> increasing 9%-29% relative to hybrid-IR. VMI at 40 keV with DLR significantly increased d' by 307% for high-density and 283% for low-density calcifications compared to conventional hybrid-IR. At 50% dose reduction, VMI at 40 keV with DLR Smooth improved CNR<sub>lumen</sub> (+ 46%), and TTF<sub>50</sub> (+ 18%) and d' (+ 24%) of the high-density calcification relative to full-dose hybrid-IR 40 keV VMI, whereas confidence intervals overlapped for NPS<sub>mean</sub> and for TTF<sub>50</sub> and d' of the low-density calcification. On dual-layer CT, spectral DLR outperforms hybrid-IR on both conventional and VMI images in noise, CNR, spatial resolution, and plaque detectability. At 50% dose reduction, DLR Smooth on 40 keV VMI exceeded full-dose hybrid-IR on 40 keV VMI in CNR, and in spatial resolution and detectability of high-density calcifications, with overlapping confidence intervals for noise texture and for low-density calcifications. Patient studies are needed to validate clinical translation.