Smart Spectral Doppler Ultrasound Imaging using Motion Tracking based on Wide-beam transmission and Neural Networks: Preliminary Results
Authors
Abstract
Ultrasound spectral Doppler imaging provides quantitative blood flow information for vascular diagnosis, but its accuracy is limited by the one-dimensional (1D) sample volume (SV). Motion from respiration, patient movement, and probe manipulation can move the target vessel out of the SV, causing spectral broadening, signal reduction, or loss of the spectrum. A two-dimensional (2D) SV can address this, but sequential focused-beam transmission reduces the pulse-repetition frequency (PRF), and plane-wave transmission lowers the signal-to-noise ratio (SNR). In this study, we propose a wide-beam transmission method that acquires a 2D SV from a single transmit event while preserving the PRF. A Hamming network, which requires no prior training, tracks axial and lateral motion within the 2D SV and updates the position of an adaptive 1D SV for each frame. Field II simulations showed that the wide-beam transmission had a spectral SNR comparable to that of focused-beam transmission and approximately 10 dB higher than that of plane-wave transmission. The motion tracking correctly detected the target position in over 93% of frames for vessel radii greater than 1 mm, and in 100% of frames for radii greater than 2 mm. In vivo studies of a carotid artery (linear array) and a hepatic vein (convex array) showed that the proposed method recovered obscured spectra and increased the spectral SNR by 9.3 dB and 4.5 dB, respectively. These results demonstrate that the proposed wide-beam transmission with Hamming network-based tracking can improve the accuracy of spectral Doppler imaging under tissue motion. Moreover, we anticipate that the proposed approach can be implemented in real-time due to its single transmission and simple two-layer architecture with low computational complexity.