Measurement-Aware Multimodal LLM for No-Reference Low-Dose CT Quality Assessment

November 5, 2026arxiv logopreprint

Authors

Kerem Basim,Tolga Demiroglu,Mehmet Ozan Unal,Metin Ertas,Isa Yildirim

Abstract

No-reference image quality assessment (IQA) of low-dose CT predicts the score that radiologists would assign to a slice without a full-dose reference, as needed to compare dose protocols and denoisers. Trained CT IQA networks agree well with radiologists but output only a score, whereas multimodal large language models (MLLMs) can explain a score but perform poorly on CT when only prompted. We propose a measurement-aware MLLM, built on Qwen3.8-27B, that rates a slice from its image together with six image-domain measurements of noise, sharpness, high-frequency content and streaks, given to the language model as percentiles in text. On the LDCT-IQA 2023 test set, it reaches an Overall score of 2.735, on par with the retrained challenge winner (2.735) and MedSigLIP-FiLM (2.713), and substantially above its base MLLM used as a zero-shot agent. Shuffling the measurements at test time reduces the score by 0.048, mostly through noise, but the same measurements also lead the model to assign higher scores to over-smoothed denoised images.

Topics

eess.IV