Intended Use

Indicated for digital imaging solution designed including head, neck, cervical spine, arm, leg and peripheral (foot, hand, wrist, fingers, etc.). Intended to replace film-based radiographic diagnostic systems and provide a case diagnosis and treatment planning for physicians and other health care professionals. Not to be used for mammography.

Technology

Digital solid-state X-ray detectors using a scintillator (Cesium Iodide) coupled to Indium Gallium Zinc Oxide (IGZO) on a TFT sensor, connected via wired LAN to a PC, and capable of converting RAW images to DICOM format. Includes updated image processing and generator integration compared to predicate device.

Performance

Both clinical and non-clinical tests, including MTF, DQE, and NPS performance, showed that 0909FCE and 0909FCE-HS are equivalent or superior to the predicate in diagnostic image quality. Software validation confirmed intended function and risk mitigation for updated features. Testing performed per IEC 62220-1, IEC 60601-1, IEC 60601-1-2, and relevant FDA guidance.

Predicate Devices

No predicate devices specified